Turb 430 IR/T

What to do if...

55

ba75507e03

07/2006

6

What to do if...

6.1

General errors

Display, LoBat

Instrument does not

react to keystroke

Error message,

Error

0, 8, 16, 16384

6.2

Turbidity

Error message,

Measured values

obviously incorrect

Measured value display

< 0.01 FNU

Cause

Remedy

– The batteries or accumulator

pack are largely depleted

– Insert new batteries

– Charge the accumulator pack

(see section 3.2)

Cause

Remedy

– Software error

– Operating condition undefined

or EMC load unallowed

– Processor reset:

Press the <START/ENTER>
and <PRT> key
simultaneously.

Cause

Remedy

– Instrument error

– Repeat measurement

– Meter defective,

send meter to WTW for repair
and quote the error number

Cause

Remedy

– Cell not correctly inserted

– Lock cell into place

– Cell contaminated

– Clean the cell

– Calibration too old

– Carry out calibration

Cause

Remedy

– Calibration defective

– Carry out calibration

– Measured value outside the

measuring range

– not possible