7 technical data, 1 general data, 1 turb430ir/t – Xylem TURB 430 PORTABLE TURBIDITY METER User Manual

Page 57: Technical data, General data 7.1.1, Turb 430 ir/t, Chapter 7 t, Echnical, Data, See chapter 7 t

Advertising
background image

Turb 430 IR/T

Technical data

57

ba75507e03

07/2006

7

Technical data

7.1

General data

7.1.1

Turb 430 IR/T

Dimensions

approx. 236 x 86 x 117 mm

Weight

approx. 0.6 kg (without batteries)

Mechanical structure

Type of protection

IP 67

Electrical safety

Protective class

III

Test certificates

cETLus, CE, FCC

Ambient

conditions

Storage

- 25 °C ... + 65 °C

Operation

0 °C ... + 50 °C

Climatic class

2

Allowable relative

humidity

Yearly mean:
30 days /year:
other days:

75 %

95 %

85 %

Power

supply

Batteries

4 x 1.5 V, type AA

Operating time with
battery operation

Turb 430 IR: approx. 3000 measurements
Turb 430 T: approx. 2000 measurements

Accumulator pack
(optional)

5 x 1.2 V nickel metal hydride (NiMH),
type AA

Power pack
Charging device
(optional)

FRIWO FW7555M/09, 15.1432.500-00
Friwo Part. No. 1883259
Input: 100 ... 240 V ~ / 50 ... 60 Hz / 400 mA
Output: 9 V = / 1.5 A
Connection max. overvoltage category II

Primary plugs contained in the scope of
delivery: Euro, US, UK and Australian.

Serial

interface

Connection of the cable AK 540/B or AK 540/S

Baud rate

adjustable:
1200, 2400, 4800, 9600, 19200 Baud

Type

RS232

Data bits

8

Stop bits

2

Parity

None

Handshake

RTS/CTS

Cable length

max. 15 m

Advertising