Turb 430 IR/T
Technical data
57
ba75507e03
07/2006
7
Technical data
7.1
General data
7.1.1
Turb 430 IR/T
Dimensions
approx. 236 x 86 x 117 mm
Weight
approx. 0.6 kg (without batteries)
Mechanical structure
Type of protection
IP 67
Electrical safety
Protective class
III
Test certificates
cETLus, CE, FCC
Ambient
conditions
Storage
- 25 °C ... + 65 °C
Operation
0 °C ... + 50 °C
Climatic class
2
Allowable relative
humidity
Yearly mean:
30 days /year:
other days:
75 %
95 %
85 %
Power
supply
Batteries
4 x 1.5 V, type AA
Operating time with
battery operation
Turb 430 IR: approx. 3000 measurements
Turb 430 T: approx. 2000 measurements
Accumulator pack
(optional)
5 x 1.2 V nickel metal hydride (NiMH),
type AA
Power pack
Charging device
(optional)
FRIWO FW7555M/09, 15.1432.500-00
Friwo Part. No. 1883259
Input: 100 ... 240 V ~ / 50 ... 60 Hz / 400 mA
Output: 9 V = / 1.5 A
Connection max. overvoltage category II
Primary plugs contained in the scope of
delivery: Euro, US, UK and Australian.
Serial
interface
Connection of the cable AK 540/B or AK 540/S
Baud rate
adjustable:
1200, 2400, 4800, 9600, 19200 Baud
Type
RS232
Data bits
8
Stop bits
2
Parity
None
Handshake
RTS/CTS
Cable length
max. 15 m