Waveform events, Waveform events -4, Table 6-1 – Physio-Control LIFEPAK 20e User Manual

Page 112

Advertising
background image

Data Management

6-4

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

Waveform Events

In addition to being documented in the Event Log, therapy and other selected events also capture
waveform data as described in

Table 6-2

.

*To reduce the length of the CODE SUMMARY report, storing waveform data with these events can be configured

OFF

(refer to

page 8-11

).

Table 6-1

Event Types

Event Types

Events

Monitoring

• Initial rhythm

• Alarm events

• Vital signs

Operator
initiated

• Event

• Print

• Sync On/Off

• Internal Pacer

Detection On/Off

• Alarms On

• VF/VT Alarm On/

Off

Therapy

AED

• AED mode

• Connect

electrodes

• Analysis

• Shock advised

• Check

patient

• Motion

• Analysis

stopped

• No shock advised

Defibrillation

• Manual

mode

• Charge removed

• Shock X

Delivered

• Shock X Not

Delivered

Pacing

• Started

• Set

• Changed

• Stopped

• Paused

Memory
Status

• Out of

Waveform
Memory
[memory low]

• Out of Event

Memory [memory
full]

Table 6-2

Waveform Events

Event Name

Waveform Data

INITIAL RHYTHM

• 8 seconds after leads on.

CHECK PATIENT

• 8 seconds prior to alert.

SHOCK or NO SHOCK
ADVISED

• 2–3 segments of analyzed ECG. Each segment may be 2.7

seconds.

ANALYSIS X STOPPED

• 8 seconds of data prior to cessation of analysis.

SHOCK X

• 3 seconds prior to shock and 5 seconds after shock.

PACING X STARTED

• 8 seconds prior to increase of current from 0.

PACING X SET

• 8 seconds after ppm and mA are stable for 10 seconds.

PACING X CHANGED

• 8 seconds after pacing rate, current, or mode is changed.

PACING X STOPPED

• 3 seconds prior to pacing current is zero and 5 seconds after.

PACING X PAUSED

• Initial 8 seconds while

PAUSE

is pressed.

ALARM*

• 3 seconds prior to violated parameter and 5 seconds after.

EVENT*

• 3 seconds prior to event selection and 5 seconds after.

PRINT

• 3 seconds prior to pressing

PRINT

and 5 seconds after.

Advertising