Specifications and Performance Characteristics
A-2
LIFEPAK 20e Defibrillator/Monitor Operating Instructions
PHYSICAL CHARACTERISTICS
Weight (maximum)
Basic defibrillator/monitor: 5.31 kg (11.7 lb)
Fully featured defibrillator/monitor (Pacing, SpO
2
, and door)
without paper or cables: 5.58 kg (12.3 lb)
QUIK-COMBO cable, add: 0.20 kg (0.43 lb)
For Standard (Hard) Paddles, add: 0.88 kg (1.95 lb)
CodeManagement Module, add: 1.63 kg (3.6 lb)
Size (maximum)
LIFEPAK 20e device
LIFEPAK 20e device with
CodeManagement Module
Height
Width
Depth
21.3 cm (8.4 in.)
25.4 cm (10.0 in.)
26.2 cm (10.3 in.)
26.2 cm (10.3 in.)
26.2 cm (10.3 in.)
29.7 cm (11.7 in.)
DISPLAY
Size (active viewing area)
115.18 mm (4.53 in.) wide x 86.38 mm (3.4 in.) high
Display Type
320 dot x 240 dot color LCD
Displays a minimum of 3.7 seconds of ECG and
alphanumerics for values, device instructions, or prompts
Option to display one additional waveform
Waveform display sweep speed: 25 mm/sec for ECG and
SpO
2
; 12.5 mm/sec for CO
2
DATA MANAGEMENT
The device captures and stores patient data, events
(including waveforms and annotations), and continuous ECG
and CO
2
waveform records in internal memory
The user can select and print reports
Report Types
Two format types of CODE SUMMARY critical event record
(short and medium)
• Initial ECG (except short format)
• Auto vital sign measurements every 5 minutes
Memory Capacity
Two full-capacity patient records that include
CODE SUMMARY critical event records - up to 100 single
waveform events
MONITOR
ECG
ECG is monitored via several cable arrangements. A 3-wire
cable is used for 3-lead ECG monitoring. A 5-wire cable is
used for 3-lead ECG plus AVR, AVL, AVF, and C. Standard
paddles or therapy electrodes (QUIK-COMBO pacing/
defibrillation/ECG electrodes or FAST-PATCH disposable
defibrillation/ECG electrodes) are used for Paddles lead
monitoring. Compatible with LIFEPAK 12 ECG and Therapy
Cables.