Basic Orientation
2-18
LIFEPAK 20e Defibrillator/Monitor Operating Instructions
When the alarms are set to
OFF
, you must press the
ALARMS
button and select
QUICK SET
to
enable the alarms.
When you press the
ALARMS
button, the following Alarms
overlay appears:
1 Select
QUICK SET
to activate the
alarms for all active parameters. The
quick set limits are set automatically
based on the patient’s current vital
sign values (refer to
). The
alarm limits default to the setting
(
WIDE
or
NARROW
) displayed on the
overlay.
2 Select
LIMITS
to change the alarm
limits to
WIDE
or
NARROW
(refer to
3 Select
SUSPEND
to turn off the
audible alarm for up to 15 minutes. If
an alarm limit is exceeded while the
alarm is silenced, the violated
parameter flashes, an alarm
message appears, but the alarm tone
remains silent.
Select
VF/VT ALARM
to turn on
continuous monitoring for ventricular
fibrillation and ventricular tachycardia
in manual mode.
A symbol appears above the primary
ECG when the alarm is on
.
Reselect
VF/VT
ALARM
to turn off
this alarm.
Note: When the VF/VT alarm is
on, you are limited to
PADDLES
lead or lead II. Refer to
.
Note: The VF/VT alarm will be
suspended when the noninvasive
pacemaker is on and when
standard paddles are attached
and
PADDLES
lead is selected.
The alarm is also suspended
when the defibrillator is charging
or is fully charged.