Troubleshooting tips for etco2, Troubleshooting tips for etco – Physio-Control LIFEPAK 20e User Manual

Page 68

Advertising
background image

Monitoring

3-22

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

Troubleshooting Tips for EtCO

2

Table 3-4

Troubleshooting Tips for EtCO

2

Observation

Possible Cause

Corrective Action

1

ALARM APNEA

message

appears and waveform is
solid line at or near zero.

No breath has been detected
for 30 seconds since last valid
breath.

• Check the patient.

FilterLine connection to device
is loose.

• Twist FilterLine connector

clockwise until firmly seated
and can no longer be
turned.

FilterLine set is disconnected
from patient or ETT.

• Check ventilation equipment

(if used) for leaks or
disconnected tubing.

2

CO2 FILTERLINE OFF

message appears and
waveform is “---”.

FilterLine set disconnected
from device.

• Connect FilterLine set to

device port.

FilterLine set not securely
connected to device.

• Twist FilterLine connector

clockwise until firmly seated
and can no longer be
turned.

3

CO2 FILTERLINE PURGING

message appears and
waveform is “---”.

FilterLine set is kinked or
clogged with fluid.

• Disconnect and then

reconnect the FilterLine set.

4

CO2 FILTERLINE BLOCKAGE

message appears and
waveform is “---”.

The message appears after 30
seconds of unsuccessful
purging. FilterLine set is kinked
or clogged.

• Disconnect and then

reconnect the FilterLine set.

• Change the FilterLine set.

5

CO2 INITIALIZING

message

appears and waveform is
“---”.

FilterLine set connected to
device while module is
initializing.

• None.

6

AUTO ZEROING

message

appears and waveform is
“---”.

Module is performing self-
maintenance.

• None.

7 EtCO

2

values are erratic.

FilterLine connection to device
is loose.

• Twist FilterLine connector

clockwise until firmly seated
and can no longer be
turned.

A leak in the FilterLine set.

• Check for connection leaks

and line leaks to patient,
and correct, if necessary.

A mechanically ventilated
patient breathes
spontaneously or patient is
talking.

• No action required.