Remote synchronization procedure, Remote synchronization procedure -20 – Physio-Control LIFEPAK 20e User Manual

Page 90

Advertising
background image

Therapy

4-20

LIFEPAK 20e Defibrillator/Monitor Operating Instructions

Remote Synchronization Procedure

The LIFEPAK 20e defibrillator/monitor

can be configured to receive an ECG source from a

remote monitor (such as a bedside ECG monitoring system) by means of the ECG/Sync
connector on the back of the defibrillator/monitor. Refer to

"Manual Mode Setup Menu,"

page 8-5

. The remote monitor must have a sync out connector and a cable must be provided to

make this connection. Refer to the LIFEPAK 20e Defibrillator/Monitor Service Manual for more
details.

Note: The LIFEPAK 20e defibrillator/monitor can be configured to remain in synchronous
mode or to return to asynchronous mode after discharge. It is important that you know how
your defibrillator is configured. Refer to

"Manual Mode Setup Menu," page 8-5

.

To perform the synchronized cardioversion using a remote monitoring ECG source:

1 Ensure

defibrillator/monitor

is connected to AC power.

2 Connect the sync cable to the defibrillator/monitor ECG/Sync connector and the remote

monitor.

3 Press

ON

.

4 Attach the ECG cable from the remote monitor to the patient.

5 Press

SYNC

on the defibrillator/monitor.

6 Select

REMOTE SYNC

from the menu.

Note: The screen on the defibrillator/monitor will display the message

REMOTE SYNC

in

place of any waveforms.

7 Observe the ECG rhythm on the remote monitor. Confirm that a sense marker appears above

each QRS complex.

8 Confirm that the sync LED on the defibrillator/monitor blinks with each detected QRS on the

remote monitor.

9 Follow Step 6 through Step 14 from

"Synchronized Cardioversion Procedure"

provided

previously.

WARNINGS!

Possible lethal arrhythmia.

Ventricular fibrillation may be induced with improper synchronization. The hospital’s biomedical
engineering staff should perform synchronization delay measurements on the system, as a
whole, to ensure that the 60 ms limit for synchronization delay is not exceeded, per requirements
as specified in AAMI DF2 (1996). Always confirm proper placement of the sense markers on the
ECG.

Possible monitor incompatibility

If the R-wave synchronization markers do not appear to be nearly simultaneous with the R-
waves on the remote monitor display or are not present, do not proceed with synchronized
cardioversion.

Advertising