If the monitor detects ECG leads off during pacing, pacing continues at a fixed rate until the ECG
lead is reattached. During fixed-rate pacing, the pacemaker delivers pulses at the set pace rate
regardless of any intrinsic beats that the patient may have. The monitor continues to display the
pacing rate (ppm) and the current (mA). To reestablish demand pacing, reattach the ECG lead.

While pacing, visually monitor the patient at all times, do not rely on the

ECG LEADS OFF

warning

to detect changes in pacing function. Routinely assess the ECG for proper sensing, pace pulse
delivery, electrical capture, and mechanical capture.

If pacing electrodes detach during pacing, the

CONNECT ELECTRODES

and

PACING STOPPED

messages appear and an alarm sounds. The pacing rate is maintained and the current resets to
0 mA. Reattaching the pacing electrodes silences the alarm and removes the

CONNECT

ELECTRODES

message. The current remains at 0 mA until you increase the current manually.

Troubleshooting Tips for Noninvasive Pacing

Table 4-4

Troubleshooting Tips for Noninvasive Pacing

Observation

Possible Cause

Corrective Action

1 Device does not function

when

PACER

is pressed.

Power off.

• Check if power is

ON

.

Low battery.

• Connect to AC power.

2

PACER

LED on, but

CURRENT (MA)

will not

increase.

Therapy electrodes off.

• Check for message

displayed.

• Inspect therapy cable and

electrode connections.

3

PACER

LED on,

CURRENT (MA) >0,

but

pace markers absent (not
pacing).

Pacing rate set below patient's
intrinsic rate.

• Increase pacing rate.

Pacer oversensing (ECG
artifact, ECG size too high).

• Establish clean ECG;

decrease ECG size.

• Select nondemand pacing.

4 Pacing stops

spontaneously.

PACER

button pressed off.

• Press

PACER

and increase

the current.

Internal error detected. Service
message indicates an internal
failure.

• Check for service indicator.
• Cycle power and start pacing

again.

• Obtain service by qualified

service personnel.

Therapy electrode off.

• Check for message. Check

pacing cable and electrode
connections.

ENERGY SELECT

or

CHARGE

pressed.

• Press

PACER

and increase

current.

• Check for service message.
• Cycle power and start pacing

again.

Radio frequency interference.

• Move radio equipment away

from pacemaker.