Clock pod characteristics – Atec Agilent-1670G User Manual

Page 14

Advertising
background image

14

CLKin

Clock Pod Characteristics

10460A TTL Clock Pod

Clock output type

10H125 with 47

Ω series; true & inverted

Clock output rate

100 MHz maximum

Clock out delay

11 ns maximum in 9 steps

Clock input type

TTL – 10H124

Clock input rate

dc to 100 MHz

Pattern input type

TTL – 10H124 (no connect is logic 1)

Clock-in to clock-out

approximately 30 ns

Pattern-in to recognition

approx. 15 ns + 1 clk period

Recommended lead set

Agilent 10474A

10463A ECL Clock Pod

Clock output type

10H116 differential unterminated; and differential with
330

Ω to –5.2V and 47 Ω series

Clock output rate

200 MHz maximum

Clock out delay

11 ns maximum in 9 steps

Clock input type

ECL – 10H116 with 50 K

Ω to –5.2V

Clock input rate

dc to 200 MHz

Pattern input type

ECL – 10H116 with 50 K

Ω (no connect is logic 0)

Clock-in to clock-out

approximately 30 ns

Pattern-in to recognition

approx. 15 ns + 1 clk period

Recommended lead set

Agilent 10474A

10468A 5 Volt PECL Clock Pod

Clock output type

10EL90 (5V) with 348

Ω pulldown to ground and 42 Ω in series

Clock output rate

300 MHz maximum

Clock out delay

11 ns maximum in 9 steps

Clock input type

100EL91 PECL (5V), no termination

Clock input rate

dc to 300 MHz

Pattern input type

100EL91 PECL (5V), no termination (no connect is logic 0)

Clock-in to clock-out

approximately 30 ns

Pattern-in to recognition

approx. 15 ns + 1 clk period

Recommended lead set

Agilent 10498A

42

348

100EL90

100EL91

CLKout

CLKin

10H125

47

WAIT

10H124

CLKout

10H116

50 k

VBB

CLKin

10H116

47

330

-5.2V

-5.2V

CLKout