Atec Agilent-1670G User Manual

Page 7

Advertising
background image

7

Agilent Technologies 1670G Series
Specifications and Characteristics

Probes (general-purpose lead set)

Input resistance

100 k

Ω, ±2%

Parasitic tip capacitance

1.5 pF

Minimum voltage swing

500 mV, peak-to-peak

Threshold accuracy*

±(100 mV + 3% of threshold setting)

Maximum input voltage

±40 V peak

State Analysis

Minimum state clock pulse width

3.5 ns

Time tag resolution

[3]

8 ns or ± 0.1% (whichever is greater)

Maximum time count
between states

34.4 seconds

Maximum state tag
count between states

[3]

4.29 x 10

9

states

Minimum master-to-master
clock time*

6.67 ns

Minimum master-to-slave
clock time

0.0 ns

Minimum slave-to-master
clock time

4.0 ns

Clock qualifier
setup/hold

4.0/0 ns fixed

Timing Analysis

Sample period accuracy

0.01% of sample period

Channel-to-channel skew

2 ns typical (not > 3 ns)

Time interval accuracy

± (sample period accuracy + channel-to-channel
skew + 0.01% of time interval reading)

Minimum detectable glitch

3.5 ns

Triggering

Sequencer speed

>150 MHz

Maximum occurrence counter

1,048,575

Range width

32 bits each

Timer value range

400 ns to 500 seconds

Timer resolution

16 ns or 0.1% (whichever is greater)

Timer accuracy

±32 ns or ±0.1% (whichever is greater)

Operating Environment

Temperature

Instrument: 0°C to 55° C (+32°F to 131°F)
Disk media: 10°C to 40°C (+50°F to 104°F)
Probe lead sets and cables:
0°C to 65°C (+32°F to 149°F)

Humidity

Instrument: up to 95% relative humidity at +40° C
Disk media and hard drive: 8% to 85% relative
humidity

Altitude

4,572 m (15,000 ft)

[3]

Time or state tags halve the acquisition memory when there are no unassigned pods.

* Warranted Specifications

370 ohms

GROUND

1.5pF

7.4pF

100 K
ohm

Figure 15. Equivalent Probe Load for the
01650-61608 General-Purpose Lead Set.

Advertising