B&K Precision 889B - Manual User Manual

Page 10

Advertising
background image

9

Z Accuracy:

As shown in table 1.

C Accuracy:

Cx

f

Zx

=

π

2

1


C

Ae

= Ae of C

f

: Test Frequency (Hz)

Cx

: Measured Capacitance Value (F)

|Zx| : Measured Impedance Value (

Ω)

Accuracy applies when Dx (measured D value)

≤ 0.1

When Dx > 0.1, multiply C

Ae

by

2

1

Dx

+


Example:

Test Condition:

Frequency

:

1KHz

Level

:

1Vrms

DUT

:

100nF

Then

=

=

=

1590

9

10

100

3

10

2

1

2

1

π

π

Cx

f

Zx

Refer to the accuracy table, get C

Ae

=±0.1%

L Accuracy:

Lx

f

Zx

=

π

2


L

Ae

= Ae of L

f

: Test Frequency (Hz)

Lx

: Measured Inductance Value (H)

|Zx| : Measured Impedance Value (

Ω)

Accuracy applies when Dx (measured D value)

≤ 0.1

When Dx > 0.1, multiply L

Ae

by

2

1

Dx

+


Example:

Test Condition:

Frequency

: 1KHz

Level

: 1Vrms

DUT

: 1mH

Then

=

=

=

283

.

6

3

10

3

10

2

2

π

π

Lx

f

Zx


Refer to the accuracy table, get L

Ae

= ±0.5%

Advertising