B&K Precision 889B - Manual User Manual
Page 10

9
Z Accuracy:
As shown in table 1.
C Accuracy:
Cx
f
Zx
⋅
⋅
⋅
=
π
2
1
C
Ae
= Ae of C
f
: Test Frequency (Hz)
Cx
: Measured Capacitance Value (F)
|Zx| : Measured Impedance Value (
Ω)
Accuracy applies when Dx (measured D value)
≤ 0.1
When Dx > 0.1, multiply C
Ae
by
2
1
Dx
+
Example:
Test Condition:
Frequency
:
1KHz
Level
:
1Vrms
DUT
:
100nF
Then
Ω
=
−
⋅
⋅
⋅
⋅
=
⋅
⋅
⋅
=
1590
9
10
100
3
10
2
1
2
1
π
π
Cx
f
Zx
Refer to the accuracy table, get C
Ae
=±0.1%
L Accuracy:
Lx
f
Zx
⋅
⋅
⋅
=
π
2
L
Ae
= Ae of L
f
: Test Frequency (Hz)
Lx
: Measured Inductance Value (H)
|Zx| : Measured Impedance Value (
Ω)
Accuracy applies when Dx (measured D value)
≤ 0.1
When Dx > 0.1, multiply L
Ae
by
2
1
Dx
+
Example:
Test Condition:
Frequency
: 1KHz
Level
: 1Vrms
DUT
: 1mH
Then
Ω
=
−
⋅
⋅
⋅
=
⋅
⋅
⋅
=
283
.
6
3
10
3
10
2
2
π
π
Lx
f
Zx
Refer to the accuracy table, get L
Ae
= ±0.5%