23

Set the measurement level according to the parameter. When setting is done the 889B will return “OK” string.

PARAMETER:

ASCII string

Numerical code

1VDC

0

1Vrms

1

250mVrms

2

50mVrms

3

Example:

LEV 1V

OK

LEV?

Return the current measurement level setting.

Example:

ASC ON

OK

LEV?

1Vrms

(return value)

ASC OFF

OK

LEV?

1

(return value)

MODE?

Query the measurement mode. If in LCR measurement mode, six fields will be returned.

1. Frequency

2. Level

3. Measurement mode

4. Unit of primary display

5. Unit of secondary display

The existence of field 5 depends on the measurement mode. For example, there’s no field 5 if the measurement

mode is DCR. The separation between fields is space (ASCII 20H).

Example:

ASC ON

OK

CPD

OK

MODE?

1KHz 1Vrms CpD uF

(return value)

ASC ON

OK

CPRP

OK

MODE?

1KHz 1Vrms CpRp uF Ohm (return value)


If in Voltage measurement mode, three fields will be returned.

1.

Measurement mode

2.

Unit of primary display