Q Accuracy:

De

Qx

De

Qx

Ae

Q

±

=

1

2


Q

Ae

= Ae of Q measurement value

Qx : Measured Quality Factor Value

De : Relative D Accuracy

Accuracy applies when

1

<

De

Qx


Example:

Test Condition:

Frequency

: 1KHz

Level

: 1Vrms

DUT

: 1mH

Then

=

=

=

283

.

6

3

10

3

10

2

2

π

π

Lx

f

Zx

Refer to the accuracy table, get

L

Ae

=±0.5%,

005

.

0

100

±

=

±

=

Ae

De

If measured Qx = 20

Then

1

.

0

1

2

1

2

±

=

±

=

De

Qx

De

Qx

Ae

Q

θ Accuracy:

100

Ae

π

180 ⋅

=

Ae

θ


Example:

Test Condition:

Frequency

: 1KHz

Level

: 1Vrms

DUT

: 100nF

Then

=

=

=

1590

9

10

100

3

10

2

1

2

1

π

π

Cx

f

Zx


Refer to the accuracy table, get

Z

Ae

=±0.1%,

deg

057

.

0

100

1

.

0

180

100

180

±

=

±

=

±

=

π

π

θ

Ae

Ae