2.2 Making Measurement

2.2.1 Open and Short Calibration

The 889B provides open/short calibration capability so the user can get better accuracy in measuring high and

low impedance. We recommend that the user perform open/short calibration if the test level or frequency has

been changed.

 Open Calibration

First, remaining the measurement terminals at the open status, press the Open key then the LCD will display:

This calibration takes about 15 seconds. After it is finished, the 889B will beep to show that the calibration is

done.

 Short Calibration

To perform the short calibration, insert the Shorting Bar into the measurement terminals. Press the Short key

then the LCD will display:

This calibration takes about 15 seconds. After it is finished, the 889B will beep to show that the calibration is

done.

2.2.2 Relative Mode

The relative mode lets the user to make a quick sort of a bunch of components. First, insert the standard value

component to get the standard value reading. (Approximately 5 seconds to get a stable reading.) Then, press the

Relative key, the primary display will reset to zero. Remove the standard value component and insert the

unknown component, the LCD will show the value that is the difference between the standard value and unknown

value.

2.2.3

Range Hold

To set the range hold, insert a standard component in that measurement range. (Approximately 5 seconds to get

a stable reading.) Then, by pressing the Range Hold key it will hold the range within 0.5 to 2 times of the current

measurement range. When the Range Hold is pressed, the LCD will display: