5-Terminal connection is the combination of 3T and 4T (Figure 4.4). It has four coaxial cables. Due to the

advantage of the 3T and 4T, this connection can widely increase the measurement range for 10m

Ω to 10MΩ.

(d) WRONG 4T CONNECTION

H

POT

DUT

(b) BLOCK DIAGRAM

(a) CONNECTION

(c) TYPICAL IMPEDANCE MEASUREMENT RANGE (£[)

5T

1m 10m 100m 1

10

1K 10K 100K 1M

100

10M

H

CUR

DUT

V

A

DUT

V

A

L

POT

L

CUR

Figure 4.4

 4-Terminal Path (4TP)

4-Terminal Path connection solves the problem that caused by the test lead inductance. 4TP uses four coaxial

cables to isolate the current path and the voltage sense cable (Figure 4.5). The return current will flow through

the coaxial cable as well as the shield. Therefore, the magnetic flux that generated by internal conductor will

cancel out the magnetic flux generated by external conductor (shield). The 4TP connection increases the

measurement range from 1m

Ω to 10MΩ.