B&K Precision 889B - Manual User Manual

Page 22

Advertising
background image

21

CsRs(?)

Serial capacitance and serial resistance measurement mode setting or querying command.

CsQ(?)

Serial capacitance and quality factor measurement mode setting or querying command.

CsD(?)

Serial capacitance and dissipation factor measurement mode setting or querying command.

LpRp(?)

Parallel inductance and parallel resistance measurement mode setting or querying command.

LpQ(?)

Parallel inductance and quality factor measurement mode setting or querying command.

LpD(?)

Parallel inductance and dissipation factor measurement mode setting or querying command.

LsRs(?)

Serial inductance and serial resistance measurement mode setting or querying command.

LsQ(?)

Serial inductance and quality factor measurement mode setting or querying command.

LsD(?)

Serial inductance and dissipation factor measurement mode setting or querying command.

RsXs(?)

Serial resistance and serial reactance measurement mode setting or querying command.

RpXp(?)

Parallel resistance and parallel reactance measurement mode setting or querying command.

ZTD(?)

Impedance and angle (Deg) measurement mode setting or querying command.

ZTR(?)

Impedance and angle (Rad) measurement mode setting or querying command.

DCV(?)

DC Voltage measurement mode setting or query command.

ACV(?)

AC Voltage measurement mode setting or query command.

DCA(?)

DC Current measurement mode setting or query command.

ACA(?)

AC Current measurement mode setting or query command.

Example:

CPD (set to Cp-D measurement mode)

OK

CPD?

0.22724 0.12840 (return values)

DCR?

5.1029

(return value)

*IDN?

Query the identity of the 889B. This command is used to identify the basic information of 889B. The return value

has four fields separated by comma (,). The total length will not greater than 100 characters. The four fields are:

1.

Manufacturer Name

2.

Model Number

3.

Serial Number

4.

Firmware Version Number

Example:

*IDN?

B&K PRECISION CORP. MODEL889B,123456789,4.096

*RST

Reset the 889B to the power on default status. The default status is:

1KHz 1Vrms CpD uF

After the 889B is reset, it will return the identity string back.

ASC

Set the format of the return value. This command sets the ASCII string return or the numerical code.

PARAMETER:

ON ASCII string

OFF Numerical code

Advertising