2.2.3 Open and Short Calibration

The Models 885 & 886 provides open/short calibration

capability so the user can get better accuracy in measuring high and
low impedance. We recommend that the user performs open/short
calibration if the test level or frequency has been changed.
 Open Calibration

First, remaining the measurement terminals with the open status,
then press the CAL key shortly (no more than two second), the
LCD will display:

This calibration takes about 10 seconds. After it is finished, the
Model 885 will beep to show that the calibration is done.

 Short Calibration

To perform the short calibration, insert the Shorting Bar into the
measurement terminals. Press the CAL key for more than two
second, the LCD will display:

This calibration takes about 10 seconds. After it is finished, the
Model 885 will beep to show that the calibration is done.