2 open/short compensation – B&K Precision 886 - Manual User Manual

Page 38

Advertising
background image

35

(a) Parastic Effect

H

CUR

H

POT

L

POT

L

CUR

Cd

Connection

Point

DUT

C

h

C

l

Ground

(b) Guard Plant reduces

Parastic Effect

H

CUR

H

POT

L

POT

L

CUR

Guard

Plant

Figure 3.6

4.2 Open/Short Compensation

For those precision impedance measuring instrument, the open and
short compensation need to be used to reduce the parasitic effect of
the test fixture. The parasitic effect of the test fixture can be treated
like the simple passive components in figure 3.7(a). When the DUT
is open, the instrument gets the conductance Yp = Gp + j

ωCp

(Figure 3.7(b)). When the DUT is short, the instrument gets the
impedance Zs = Rs + j

ωLs (Figure 3.7(c)). After the open and short

compensation, Yp and Zs are for calculating the real Zdut (Figure
3.7(d)).