4. Application

4.1 Test Leads Connection

Auto balancing bridge has four terminals (H

CUR

, H

POT

, L

CUR

and

L

POT

) to connect to the device under test (DUT). It is important to

understand what connection method will affect the measurement
accuracy.

 2-Terminal (2T)

2-Terminal is the easiest way to connect the DUT, but it contents
many errors which are the inductor and resistor as well as the
parasitic capacitor of the test leads (Figure 3.1). Due to these
errors in measurement, the effective impedance measurement
range will be limited at 100

Ω to 10KΩ.

R

H

CUR

H

POT

DUT

(b) BLOCK DIAGRAM

DUT

V

A

Co

o

L

o

R

o

L

o

(a) CONNECTION

(c) TYPICAL IMPEDANCE MEASUREMENT RANGE(£[)

2T

1m 10m 100m 1

10

1K 10K 100K 1M

100

10M

L

POT

L

CUR

Figure 3.1

 3-Terminal (3T)