34

measurement range from 1m

Ω to 10MΩ.

(b) BLOCK DIAGRAM

(a) CONNECTION

DUT

V

A

(c) TYPICAL IMPEDANCE

MEASUREMENT RANGE(£[)

4T

1m 10m 100m 1

10

1K 10K 100K 1M

100

10M

H

POT

DUT

H

CUR

L

CUR

L

POT

H

POT

DUT

H

CUR

L

CUR

L

POT

(d) 4T CONNECTION WITH SHILDING

Figure 3.5

 Eliminating the Effect of the Parasitic Capacitor

When measuring the high impedance component (i.e. low
capacitor), the parasitic capacitor becomes an important issue
(Figure 3.6). In figure 3.6(a), the parasitic capacitor Cd is
paralleled to DUT as well as the Ci and Ch. To correct this
problem, add a guard plane (Figure 3.6(b)) in between H and L
terminals to break the Cd. If the guard plane is connected to
instrument guard, the effect of Ci and Ch will be removed.