ST75/ST75V MASS FLOW

FLUID COMPONENTS INTERNATIONAL LLC

This page is subject to proprietary rights statement on last page

13

Doc. No. 06EN003368 Rev. C

Calibration Parameters Verification
The instrument uses a set of predetermined calibration parameters to process flow signals. Most of these parameters should not
change. A data package located with this manual contains the “ST75/ST75V Delta R Data Sheet”. This contains the calibration
parameters stored in the flow transmitter at the factory. To verify that these parameters have not changed, complete the following:

1. Identify the appropriate Delta R Data sheets by serial number of the instrument.
2. Press [D] [ENTER] to examine each of the parameters. The [ENTER] key allows scrolling one message at a time. Use Table

4 to verify parameters with the Delta R Data sheet ST75/ST75V Parameters.

Table 4. Diagnostic Test Sequence on Display

S/W Vers ion:

dR M in:

T SpanIDA C 0:

Flow Factor:

dR M ax:

T ZeroIDA C 0:

Cmin Flow:

Cal Ref:

T SpanIDA C 1:

Cmax Flow:

Tcslp:

T ZeroIDA C 1:

Eng Units:

Tcslp 0:

State 0:

Line Size 0:

Tcslp 2:

Switch Pt 0:

Line Size 1:

Tot M enu:

State 1:

Cmin Temp:

Tot Flag:

Switch Pt 1:

Cmax Temp:

Totalizer:

K factor 1:

M in Flow:

Rollover Cnt:

K factor 2:

M ax Flow:

Fix Pt Flag:

K factor 3:

Dens ity:

Puls e Factor:

K factor 4:

*C1 [1]:

Puls e Out:

I factor:

*C1 [2]:

Hours:

Temp Flag:

*C1 [3]:

Sample Period:

Out M ode:

*C1 [4]:

dR Slope:

Boxcar Max:

*C1 [5]:

dr Off Set:

RTD-SLP-385:

Break Pt:

Refr Slope:

% of Range:

*C2 [1]:

Refr Off Set:

Us er Name:

*C2 [2]:

SpanIDA C 0:

Shop Order #:

*C2 [3]:

ZeroIDA C 0:

Serial No.:

*C2 [4]:

SpanIDA C 1:

M odel#:

*C2 [5]:

ZeroIDA C 1:

If parameters that have an asterisk (*) have changed, this may indicate a problem. Customer Service should be contacted. If the
parameters have not changed, continue with the next section.