ST75/ST75V MASS FLOW

FLUID COMPONENTS INTERNATIONAL LLC

This page is subject to proprietary rights statement on last page

9

Doc. No. 06EN003368 Rev. C

Example: COMMAND V (Reference Table 3)

Case: 4-20mA #1 = flow, 4-20mA #2 = Temperature,

Source Out = Pulse,

Sink = Alarm

Pressing [V] [ENTER] will display

Output Mode Selected”

followed by:

“4-20mA #1 = Flow”

“ 4-20mA #2 = Temp”

followed by

“Change? (Y/N)”
Press [ENTER]
(no change).

The last saved mode will display at this point. i.e.,

“Source: Pulse”

“Sink: Pulse”

followed by,

“Change? (Y/N)”

Select Y [Enter].

The display reads,

“Source: Pulse”

“Sink: Pulse”

followed by,

“Enter 1 to make the selction #.”

Select [ENTER].

The next display reads,

“Source: Pulse”

“Sink: Alarm”

followed by,

“Enter 2 to make the selction #.”

Select 2 and [ENTER].

The next prompt reads,

“PFactor: 1.000”

“Change? (Y/N)>”

(this factor can be anywhere from 0.001 to 1000 - A
pulse factor of 1.000 will output 1 pulse per unit of
flow.)

If no change, select N and/or [ENTER] to continue.

The next prompt is, “Sample Period”
“Change? (Y/N)>”

(this value may be set from 0.5 to 5 seconds)

If no change, select N and/or [ENTER] to continue.

The next prompt is, “Source state:

“High to Low” Change to “Low to High?>”

(this selection toggles the pulse
signal normally high or normally
low).

[ENTER] to read display.
“Switchpt1”

“0.0000000”

the current set point.

“Change? (Y/N)>” enter Y [ENTER] and enter #______ . Set Point Value , i.e. 50 (value is in same units as the flow and
must be within the calibrated range). [ENTER]. The next prompt is,
Sink state:

“High to Low” Change to “Low to High?>”. Set the output signal to be normally “High” or

normally “Low.” Pressing [Y]

[ENTER] toggles the current setting. Pressing [ENTER] resumes normal operation.

NOTE: The display comes up to the last setting saved and stays for 2 seconds. If N or [ENTER] is entered, the menu
proceeds to the Pulse out. If Y is entered, the display moves to the selection options and/or asks for confirmation. If you
miss the option, select [Enter] repeatedly to loop around.

“V” Menu Output Configuration Set Up

Analog out
Output Mode
Selected

4-20mA #1:

Flow

4-20mA #2: Temp

Change? (Y/N)>

4-20mA #1:

Flow

4-20mA #2:

Temp
Enter 1 to make
the selection__

4-20mA #1:
4-20mA #2:

Flow
Enter 2 to make
the selection__

4-20mA #1:

Temp

4-20mA #2:

Flow
Enter 3 to make
the selection__

4-20mA #1:

Temp

4-20mA #2:

Temp
Enter 4 to make
the selection__

Pulse out
Pulse Out
Selected

Source:

Pulse

Sink:

Pulse

Change? (Y/N)>

Source:

Pulse

Sink:

Pulse
Enter 1 to make
the selection #__

Source:

Pulse

Sink:

Alarm1
Enter 2 to make
the selection #__

Source:

Alarm0

Sink:

Pulse
Enter 3 to make
the selection #__

Source:

Alarm0

Sink:

Alarm1
Enter 4 to make
the selection #__

PFactor: 1.000
Change? (Y/N)>

if yes
Enter new factor: ____

Sample Period: 1 second

Change? (Y/N)>
if yes
Enter new Sample Period: ____

If alarm is a selected output
Set point1: 000

Set points are in the
same units as the flow
or temp.

Change? (Y/N)>

if yes
Enter new set point: ____

Resume normal operation

Source state:
High to Low

Change to Low to High?>