Specifications

Media Compatibility:

Air, compressed air, nitrogen, oxygen, argon, CO2, ozone,

other inert gases, natural gas, and other hydrocarbon

gases, hydrogen.

Pipe/Line Size Compatibility: 1/4" to 2" [6 mm to 51 mm]

Instrument

ST75/ST75V Range:*

* actual range subject to gas type and specific conditions.

Accuracy:

ST75: ± 2% of reading, ± 0.5% of full scale

Optional: ± 1% of reading, ± 0.5% of full scale

ST75V: ± 1% of reading, ± 0.5% of full scale

Repeatability: ± 0.5% of reading

Temperature Compensation:

Standard: 40 to 100°F [4 to 38°C]

Optional: 0 to 250°F [-18 to 121°C]

Turndown Ratio: 10:1 to 100:1

Agency Approvals:

ATEX/IEC Ex: II 2 G Ex d IIC T6...T3

II 2 D Ex tD A21 IP67 T90°C...T121°C

FM, CSA:

Class I, II, III, Div. 1, Groups B-G

Class I, Div. 2, Groups A-D

CRN No.:

0F0303

Warranty: One year

Flow Element

Installation:

ST75: In-line “T”, NPT or tube

Material of Construction: All-welded 316 stainless steel

probe element with Hastelloy-C thermowells; 316 stainless

steel NPT and tube fittings

Maximum Operating Pressure:

ST75:

T-fitting [NPT female]: 240 psi [16.5 bar(g)]

Tube: 600 psi [41 barg]

ST75V: 240 psi [16.5 bar(g)]

Temperature: Operating: 0 to 250°F [-18 to 121°C]

Process Connection:

ST75: T-fitting [NPT female]: 1/4”, 1/2”, 3/4”, 1”, 1 1/2” or 2”

Tubing: 1/4”, 1/2” or 1”

ST75V: Female NPT, Male NPT, Flange

1/4”, 1/2”, 3/4”, 1”, 1 1/2” or 2”

Transmitter

Enclosure: NEMA 4X [IP67], aluminum, dual conduit ports

with either 1/2" Female NPT or M20x1.5 entries. Epoxy coated.

Output Signals:

(2) 4-20 mA user assignable to flow rate and/or temperature

(1) 0-1 kHz pulse for total flow

Communication Port: RS232C standard. Optional wireless IR

to PDA with digital display models

Input Power:

DC: 18 Vdc to 36 Vdc (6 watts max.)

AC: 85 Vac to 265 Vac (12 watts max.; CE Mark Approval for 100

Vac to 240 Vac)

Operating Temperature Range: 0 to 140°F [-18 to 60°C]

Digital Display: (Optional)

Two-line x 16 characters LCD. Displays measured value and
engineering units. Top line assigned to flow rate. Second line is
user assignable to temperature reading, flow totalizer or
alternating. Display can be rotated in 90° increments for
optimum viewing orientation.

NPT

Line

Size

¼” ½” ¾” 1” 1

½” 2”

Min.

SCFM 0.04 0.13 0.22 0.35 0.85 1.40

Min

[NCMH] [0.07] [0.22] [0.38] [0.59] [1.44] [2.38]

Max.

SCFM 17.34 50.64 88.88 139.95 539.31 559.27

Max.

[NCMH] [29.47] [86.04] [151.00] [237.78] [576.48] [950.20]

Tubing Line Size

¼”

½”

1”

Min. SCFM

0.01

0.05

0.25

Min [NCMH]

[0.01]

[0.09]

[0.42]

Max.

SCFM

3.02 21.15 99.08

Max. [NCMH]

[5.14]

[35.94]

[168.33]

Advertising