Start up and commissioning – Fluid Components International ST75_ST75V User Manual

Page 7

Advertising
background image

ST75/ST75V MASS FLOW

FLUID COMPONENTS INTERNATIONAL LLC

This page is subject to proprietary rights statement on last page

7

Doc. No. 06EN003368 Rev. C

If the instrument is installed, and the process flow is zero, the instrument should indicate 0.000.

Flow Unit Modification
Example: SCFM Flow Units and 3 inch Sch 40 round pipe size set up:

Start up and Commissioning

1. Verify all Input power and output signal wiring is correct and ready for initial power start up.
2. Apply power to instrument. The instrument will initialize in the Normal Operation Mode. All outputs will be active and instru-

ments with the display option will indicate flow with the factory set flow unit. Allow 10 minutes for the instrument to warm up
and come to the thermal equilibrium.

The following FC88 commands are typical commands that are used during start up and commissioning:

Command

Name

Description

T

Normal Operation Mode

All outputs are active

Z

Flow Unit Set-Up

Select Flow Units(4 English, 4
Metric)Pipe Dimensions

W

Totalizer

Enable/Disable

V

Output Configuration

Select one of 4 Configurations:
Pulse and/or AlarmPulse factor and/or
setpoint

F

K-Factor (default=1)

Flow factor

N

Warm Re-set

Re-initialize C/B

S

Totalizer Menu

Enables W menu (Option)

Enter

Display

Description

Enter

menu: >

From Normal Operation Mode

Z

E for English M for Metric >

Flow Unit Set-Up menu

E

0=SFPS, 1=SCFM, 2=SCFH, 3=LB/H,
4=GPM #

English units

1

R round duct or S rectangular>

Select Standard Ft

3

/Min (SCFM)

Dia.: 4.0260000
Change? (Y/N)>

Y

Enter value: #
area: 7.3926572 CMinflow: 0.0000000
Change? (Y/N)>
Maximum flow: 462.04
Enter to continue
Cmaxflow: 462.04
Change? (Y/N)>

Y

#
CMintemp (F): -40.00000
Change? (Y/N)>
CMaxtemp (F): 250.00000
Change? (Y/N)>
Percent of Range is: OFF
Change to ON?>
LCD Mult Factor x1
Change? (Y/N)>

N

100.0 SCFM

Instrument will end up in Normal
Operation Mode

R

Select Round Duct

3.068

3 inch Sch. 40 pipe I.D.

N

Y

462.04

N

N

N