ST75/ST75V MASS FLOW

FLUID COMPONENTS INTERNATIONAL LLC

This page is subject to proprietary rights statement on last page

5

Doc. No. 06EN003368 Rev. C

Pulse Output Activation

The ST75/ST75V provides a pulse output feature. Instruments ordered with volumetric or mass flow units will be factory set with
totalizer and pulse output activated. The mode can be changed in the field. Wiring either sink or source mode is shown in Figures E
and F below. Though only one configuration is shown with the VAC and VDC power supplies, the source or sink can be utilized with
either power input.
Sink Mode: 40 VDC Max, 150 mA max. Customer supplied power source
Source Mode: 15 VDC output, 50 mA max

Figure E

VDC Power Connection

Figure F

VAC Power Connection

RS 232 Connection

VAC Power

As Shown:

85-265 VAC power connected with gnd
4-20mA connected for flow and temperature
Pulse Out in sink mode

Note: In sink mode, 40VDC max, 150mA max customer
supplied power source.

Power Disspation

AC Version
Power dissipation values under nominal conditions:
Instrument (Electronics + Sensor): 11.6 Watts
Sensor only: 0.25 Watts

Power dissipation values under maximum load conditions:
Instrument (Electronics + Sensor): 12 Watts

Sensor only: 0.30 Watts

DC Version
Power dissipation values under nominal conditions:
Instrument (Electronics + Sensor): 4.5 Watts
Sensor only: 0.25 Watts

Power dissipation values under maximum load conditions:
Instrument (Electronics + Sensor): 6 Watts

Sensor only: 0.30 Watts

RS 232 Connection

VDC Power

As Shown:

18-36VDC power connected with gnd
4-20mA connected for flow and temperature
Pulse Out in source mode

Note: In source mode, 15VDC Output max, 50mA max.