Routine Maintenance

Low Flow Message

The low flow message can be caused by clogged or dirty filters, excessive

water moisture in the filters, or an extremely cold pump assembly. If you get

a low flow indication, follow the directions below to clear the message.

If the Low Flow message goes away when the sample hose is disconnected,

you have excessive water in the filter system, or the filters are dirty. Run

the extended Purge a couple of times. Make sure there is no water in the

sample probe handle assembly, or in the water separator.

If the Low Flow message still does not go away, disconnect the in-line

filter that is connected directly to the analyzer, and reconnect the sample

hose to the analyzer. This will take the in-line filter out of the sample hose

circuit. If the Low Flow message goes away, replace the in-line filter. If

the Low Flow message is still on, reconnect the in-line filter and replace

the water separator filters.

Purging

During normal operation the bottom filter will be come damp and the upper

filter should remain dry. Providing the bottom filter is damp enough, purging

will clear the hose, upper filter and bowl of water. Purging will not clear

the bottom filter and sample probe handle.

Filter Replacement

The in-line filter can be replaced by simply disconnecting it from the

analyzer and fitting the replacement. The upper and lower filter in the

water separator bowl need to be replaced as a unit. Do not lose the rubber

separator that goes between the filter elements, without it, no filtering will

occur. The filter with the Red Dot goes on the bottom.

Cleaning the Sample Hose

If it becomes necessary to clean the sample hose and handle, use

clean, dry compressed air. Do not blow compressed air through the filter

elements!

How to Extend the life of the NOx and O2 Sensor

The best way to extend the life of your NOx and O2 sensor is to simply

use the analyzer. Leaving it plugged in continually does not extend the

life. The units that are experiencing longer sensor life are used several

time a week. Usage of less than once per month will shorten the life of

the sensors.