LANGUAGE (DE, EN, FR, NL)
STANDARD (concentration of the standard solution used: 1.95–2.05 in 0.01 increments)

SPECIAL P (Total P)

FACTOR (correction factor for the total phosphorus measurement values
0.20–5.00 in 0.01 increments)
CURRENT RANGE (measurement range for current output 1
0.5–5.0 in 0.1 increments)
CURRENT (current transfer for current output 1 0–20 mA, 4–20 mA)
MALFUNCTION (current transfer in the event of a fault with current output 1 off
(= last measurement value), 0 mA, 20 mA)

SPECIAL OP (Ortho-P)

FACTOR (correction factor for the orthophosphate measurement values
0.05–1.0–20.0)
CURRENT RANGE (measurement range for current output 2
0.5–5.0 in 0.1 increments)
CURRENT (current transfer for current output 2 0–20 mA, 4–20 mA)
MALFUNCTION (current transfer in the event of a fault with current output 2 off
(= last measurement value), 0 mA, 20 mA)

P/OP (number of measurement cycles per parameter each 0–10)
RELAY-MIN (lower limit-value contact 0.00 to [CURRENT RANGE P] setting)
RELAY-MAX (upper limit-value contact 0.00 to [CURRENT RANGE P] setting)
SIGMATAX (sample feed via SIGMATAX: yes, no)
DATE (setting of the date)
TIME (setting of the time)
CLEANING (automatic cleaning off, 6 h, 12 h, 1–7D, start time for each hourly --> last Display of last cleaning)
CALIBRATION (automatic calibration off, 1–7D, start time for each hourly --> last Display of last calibration)
VERSION (program version)
PASSWORD (password protection on activation 4-digits using digits 1-4)
SERVICE STATION (place and contact partner)
BACKLIGHT (display lighting)

SETTINGS

Setting of all device parameters: This includes measurement range, correction factor, current output, date, time etc.