Measurement device technical data

Subject to changes.

Performance specifications

Measurement method

Reduction method, derived from DIN 38405 D11,
analysis photometric via IR LED photometer

Measurement ranges

0.01–0.50...5.00 mg/L P total phosphorus in 0.1 mg/L increments
0.01–0.50...5.00 mg/L P orthophosphate in 0.1 mg/L increments
(depending on design)

Measurement uncertainty

0.01 – 5.00 mg/L: ±2 % from the measurement value

Measurement interval

t

100

= 10 min

Sample requirement

approximately 100 mL/h

Ambient temperature

+ 5° C to + 40° C

Calibration

Automatic at selectable intervals

Inspection interval

3 months

Maintenance requirement

approximately 1/2 hour per week, typical

Reagent stock

3 months (standard solution 6–12 months)

Outputs

2 current outputs:
2 limit-value contacts:
Service interface:
Interface:

0/4...20 mA, maximum 500 Ω
potential-free, 24 V 1A
RS 232
bus-capable (optional)

Protection type

in accordance with IP 54

Mains connection

230 V AC / 50 Hz

Power consumption

Approximately 150 VA

Dimensions: W x H x D

560 mm x 810 mm x 330 mm

Mass

Approximately 29 kg (without reagents)

Fuses

T2.5 A L; 250 V (2 units)

Other

Automated cleaning, data logger, graphic display with load curve display