English 106

Note: Add the additional components A and D to reagent solutions A
and D shortly before use (figure 5). Cool the ready prepared reagent
solutions in the refrigerator. The ready prepared reagent solutions are
usable for 3 months, after which the canisters must be disposed of
together with the remaining contents.
1.

Fill the additional component A into the reagent canister A using

the funnel.

2.

Fill the additional component D into the reagent canister D using

the funnel.

3.

Seal reagent canisters A and D with new lids and shake thoroughly

for 2 x 1 minute.

4.

Stick a label with the preparation data on each reagent canister A

and D.

5.

Press a menu key for 3 seconds and select the +SERVICE menu.

6.

Open the refrigerator and pull the reagent hoses out of the

canisters.

7.

Take the old reagent canisters from the refrigerator, unscrew the

lids and replace with the lids of the new canisters.

8.

Put new canisters in the refrigerator, insert the hoses again, close

the refrigerator, select PRE PUMPING and then CALIBRATION.

Exchange of the standard solution

The standard solution is ready to use. Exchange the standard solution
depending on consumption every 6–12 months.
1.

Remove old bottle with standard solution, unscrew lid and replace

with the lid of the new bottle.

2.

Insert new bottle, screw on the lid with the suction hose.

3.

The exact concentration of the standard solution is printed on the

bottle.

4.

Enter standard concentration in the +SETTINGS > STANDARD

menu.

Cleaning tasks

1.

Press one of the function keys F1–F4 for 3 seconds

2.

Select [+SERVICE] menu

3.

Remove hoses from all reagent canisters and bottles

4.

Remove contaminated hoses — this can be done without a

problem and without tools through the easy plug-in system (if

necessary observe

section Analysis part, page 93

)

5.

Remove contamination with suitable cleaning agents (diluted

hydrochloric acid, chlorine bleaching lye). Please absolutely

observe safety notes on the cleaning agents used!

6.

Insert the cleaned hoses again

7.

Select [PRE PUMP. & CALIB.] function to fill all hoses automatically

and to start a calibration

Reaction principle

Orthophosphate ions *) in acidic solution with molybdate ions in the
presence of antimony ions form a complex that is reduced through
ascorbic acid to phosphorus molybdenum blue. The color intensity is
proportional to the orthophosphate content *) of the sample in the
specified measurement range.
Polyphosphates and some organic phosphorus compounds are
hydrolyzed through boiling in strongly acidic solution into
orthophosphates *).
Stable phosphorus compounds are digested by sodium peroxide sulfate.
In the PHOSPHAX Σ sigma work is performed at temperatures way over
100 °C under pressure. Only under these extreme conditions are such
short digestion times achieved.