Figure 6 Routing of optical-fiber cable SIGMATAX 2 and signal lines

Terminal assignment — for current outputs and

limit-value contacts

Figure 7 View of the terminal board

In order to connect the transmission lines for current

output, limit-value contacts and bus interface, the following

work steps are required:

1.

Disconnect device from the mains supply.

2.

Open device door.

3.

Loosen the knurled screw at the top right and swivel the front panel

out forwards. The connections are located on the right on the

terminal board.

4.

Loosen the PG screw connection for the corresponding connection

cable and remove the blanking disc.

5.

Plug the cable through the screw connection into the device and

route forwards to the connection terminals.

6.

Plug cable ends into the corresponding connection terminals (wire

thickness maximum 1.5 mm

2

). The assignment of the connection

terminals is seen from

Figure 7

the view of the terminal board.

N O T I C E

Only shielded lines may be connected!

Optical-fiber cable from
SIGMATAX 2

Loosen
knurled screw

Front panel

Current output 1
Current output 2
Limit-value contacts
Interface
Power plug

Socket with
ground contact

A B A B
1 1 2 2

1

2

3

4

5

6

7

8

9

10

11

12 13 14

RS 485

32 23 32 23

Relay 1 Relay 2 Current 1 Current2

Current I = current output 1 (submenu (+SPECIAL P) or (+CURRENT 1))

Current II = current output 2 (submenu (+SPECIAL OP) or (+CURRENT 2))

Terminals for
shielding

Terminal board

Relay I = relay-min (lower limit-value contact)

Relay II = relay-max (upper limit-value contact)

(Figure shows relays in idle state.)

Interface:

bus-capable