Error messages – Hach-Lange PHOSPHAX sigma Basic User Manual User Manual

Page 109

Advertising
background image

English 109

Error messages

Error Cause

Measures

NO TUNING

Electronic error

Call customer service

TEST SETTINGS

Possible device adjustment

Call up settings and check completely, with frequent occurrence
inform customer service

SECURITY PLATE

Safety panel loose

Check attachment of the safety panel

HUMIDITY

Moisture sensor responds

Remedy cause (leak), dry off moisture sensor and acknowledge error in [Status]
menu

CU NO REACTION

Electronic error

Call customer service

CU TIMEOUT

PROCESS-TIMELIMIT

POS. ERROR

Undefined piston valve position

Check piston valve and acknowledge error in [STATUS] menu

SIGMATAX SIGNAL

SIGMATAX error

Call up [STATUS] menu and determine error cause

SIGMATAX OFF

Switched off SIGMATAX

Check fiber optic and SIGMATAX

LEVEL LIMITS

Photometer error

After acknowledgement in [STATUS] menu and another occurrence call customer
service

LEVEL OFFSET

TEMPERATURE SENSOR

Temperature measurement defective

Call customer service

AMPLIFIER Photometer

error

CALIBRATION FACTOR

Check of the dosing and of the reagents

INTERNAL BUS ERROR

Electronic error

Call customer service