0675 mV
Code Display
Action
Choices
Co.
Enter polarization voltage
Select using , edit number
using , press enter to proceed.
Select pressure unit
Select using arrow key
Press enter to proceed.
bar
(kPa, PSi)
Process pressure correction
Enter process pressure. This
value is used to correct oxygen
saturation. It has no influence
on concentration measurement
(Conc).
Select position using key and
edit number using key.
Press enter to proceed.
1.013 bar
Enter salinity correction
Select position using
and edit number using .
Press enter to proceed.
00.00
ppt*
* ppt (parts per thousand) - corresponds to g/kg
47