Measurement, Temp probe adjustment – Hach-Lange SI794 D tr User Manual User Manual

Page 74

Advertising
background image

Temp Probe Adjustment

Display

Remark

Measurement

In the measuring mode the main display shows the

configured process variable (%, mg/L, or ppm) and the

lower display the temperature.

During calibration you can return to measuring mode

by pressing the cal key, during configuration by

pressing the conf key.

(Waiting time for signal stabilization approx. 20 sec).

Display

Action

Remark

Measure the temperature of

the process medium using an

external thermometer.

Enter the measured value:

Select position using ,

edit number using ,

press enter to proceed.

Press enter to end adjustment.

HOLD will be deactivated after

20 sec.

Select calibration

(Press cal key).

Enter passcode: 1015

Select position using ,

edit number using ,

press enter to proceed.

Ready for calibration

Wrong settings change

the measurement

properties!

If an invalid code is

entered, the transmitter

returns to measuring

mode.

Default:

Current value of

secondary display

The transmitter is in the

Hold mode (display for

approx. 3 sec)