Calibration

Calibration adjusts the transmitter to the sensor.

Activation

Activate with cal

Enter passcode.

Edit parameter using  and ,

confirm/proceed using enter.

(To exit, press cal, then enter.)

HOLD

During cali-

bration the

transmitter

remains in

the Hold

mode.

During calibration the transmit-

ter remains in the Hold mode

for reasons of safety. Output

current is frozen (last value or

preset fixed value, depending

on configuration), limit and

alarm contacts are inactive. The

controller is in the configured

state, Sensoface is off, “Calibra-

tion” mode indicator is on. The

red LED blinks when “HOLD ON”

has been set.

HOLD icon

Input errors

The calibration parameters are

checked during the input. In the

case of an incorrect input “Err” is

displayed for approx. 3 sec. The

incorrect parameters cannot be

saved. Input must be repeated.

End

End with cal. The measured val-

ue is shown in the main display

alternately with “Hold”, “enter”

blinks, Sensoface is active. Press

enter key to end the Hold mode.

The measured value is displayed.

The output current remains

frozen for another 20 sec (HOLD

icon on, “hourglass” blinks).

64