Code display action choices l2. hi, N/c (n/o), 0010 sec – Hach-Lange SI794 D tr User Manual User Manual
Page 53
Advertising

Code Display
Action
Choices
L2.
Hi
(Lo)
For limit 2 function principle,
see Fig. below. Select using ,
press enter to proceed.
Limit 2 contact response
N/C: normally closed contact
N/O: normally open contact
Select using , press enter to
proceed.
N/C
(N/O)
Limit 2 setpoint
Select using , edit number using
, press enter to proceed.
0500 %
(
xxxx %)
Limit 2 hysteresis
Select using , edit number using
, press enter to proceed.
0001 %
(
xxxx %)
Limit 2 delay
The contact is activated with delay
(deactivated without delay)
Select position using ,
edit number using ,
press enter to proceed.
0010 sec
(
0 ... 600 sec)
Limit Lo
Hysteresis +
Setpoint
Contact
0
1
Signal
Limit Hi
Hysteresis -
Signal
Setpoint
Contact
0
1