Error messages, Error codes), Error display problem possible causes – Hach-Lange SI794 D tr User Manual User Manual
Page 82
Advertising

Alarm contact
Red LED
Out 1 (22 mA)
Out 2 (22 mA)
Error messages
(error codes)
SAT range
Sensor defective
Wrong sensor connected
Measuring range exceeded
Error
Display Problem
Possible causes
ERR 01
Measured
value
blinks
x
x
x
Conc range
Sensor defective
Wrong sensor connected
Measuring range exceeded
ERR 02
Measured
value
blinks
x
x
x
System error
Configuration or calibration data
defective; completely reconfigure
and recalibrate the transmitter.
Memory error in device program
(PROM defective)
ERR 98
“ConF”
blinks
x
x
x
x
Factory settings
EEPROM or RAM defective
This error message only occurs
in the case of a total defect. The
transmitter must be repaired and
recalibrated at the factory.
ERR 99
“FAIL”
blinks
x
x
x
x
82