Installation

7. Repeat steps to add additional probes/devices and tags.

8. Press the

ENTER

button to save the configuration.

Table 1 Telegram list—Column description

Column

Description

1

Data position in the configured Profibus slave (in 2 byte words)
Modbus: Data position in the configured Modbus slave
This slave contains holding registers beginning at 40001.
Example: "0" means register 40001 and "11" means register 40012.

2

Tag name to identify the configured data.

3

Data type
float=floating point value
int=integer values
sel=integer value resulting from an enumeration or selection list

4

Data status
r=data is read only
r/w=read/write

Table 2 Error register

Bit

Error

Description

0

Calibration error

Faulty calibration detected

1

Electronic settings error

Faulty electronic calibration/settings

2

Cleaning error

Error in cleaning cycle detected

3

Measuring module error

Error in measuring module detected

4

System initialization

Inconsistent settings detected, reset to factory
settings

5

Hardware error

Faulty hardware detected

6

Internal communication error

Internal communication error detected

7

Humidity error

Excessive humidity detected

8

Excessive temperature

Excessive temperature detected

9
10

Sample feed warning

Error in sample feed detected

11

Questionable calibration warning

Accuracy of previous calibration inadequate

12

Questionable measurement
warning

Accuracy of previous measurement inadequate/out of
range

13

Safety warning

Safety equipment error detected

14

Reagent warning

Reagent warning, e.g. fill level < min detected

15

Service request warning

Service request detected