Figure 5, Figure 6 – Hach-Lange ORBISPHERE 410 Basic User Manual User Manual

Page 11

Advertising
background image

Figure 5 EC measurement board

Figure 6 TC measurement board

Figure 7 Connector J7

Figure 8 Connector J8

Connector J7 (inputs & outputs)
The numbers listed below refer to the 16 available J7 connections (from left to right) in

Figure 7

.

Measurement alarms relays:
1. Common
2. Output relay 1
3. Output relay 2
4. Output relay 3

Analog current (or voltage) outputs:
5. GND
6. Output 1
7. Output 2
8. Output 3

Digital inputs:
9. EC sensor: Not used
9. TC sensor: Hold input (short circuit to pin 12)
10. to 16. Not used

Connector J8 (sensor)
The numbers listed below refer to the 10 available J8 connections (from left to right) in

Figure 8

. The

colors indicated are the wire colors in the sensor cable.

Note: Remember, this connector is colored orange for EC sensors and black for TC sensors.

A1100 EC sensor

31xxx EC sensor

31xxxS smart EC
sensor

TC sensor

Sensor
cable

1.

Guard electrode

Guard electrode

Guard electrode

GND for power

Yellow

2.

RS485A+

Not used

I2C-SCL

V2 signal

Pink

3.

Thermistor A

Thermistor A

Thermistor A

Solenoid

Grey

English 11