Figure 5, Figure 6 – Hach-Lange ORBISPHERE 410 Basic User Manual User Manual
Page 11

Figure 5 EC measurement board
Figure 6 TC measurement board
Figure 7 Connector J7
Figure 8 Connector J8
Connector J7 (inputs & outputs)
The numbers listed below refer to the 16 available J7 connections (from left to right) in
.
Measurement alarms relays:
1. Common
2. Output relay 1
3. Output relay 2
4. Output relay 3
Analog current (or voltage) outputs:
5. GND
6. Output 1
7. Output 2
8. Output 3
Digital inputs:
9. EC sensor: Not used
9. TC sensor: Hold input (short circuit to pin 12)
10. to 16. Not used
Connector J8 (sensor)
The numbers listed below refer to the 10 available J8 connections (from left to right) in
. The
colors indicated are the wire colors in the sensor cable.
Note: Remember, this connector is colored orange for EC sensors and black for TC sensors.
A1100 EC sensor
31xxx EC sensor
31xxxS smart EC
sensor
TC sensor
Sensor
cable
1.
Guard electrode
Guard electrode
Guard electrode
GND for power
Yellow
2.
RS485A+
Not used
I2C-SCL
V2 signal
Pink
3.
Thermistor A
Thermistor A
Thermistor A
Solenoid
Grey
English 11