Measurement alarm relays, Sensor installation, Ec sensors – Hach-Lange ORBISPHERE 410 Basic User Manual User Manual

Page 12: Tc sensors, User interface, Instrument controls

Advertising
background image

A1100 EC sensor

31xxx EC sensor

31xxxS smart EC
sensor

TC sensor

Sensor
cable

4.

Anode electrode

Anode electrode

Anode electrode

Relay coil

Red

5.

RS485B

Not used

I2C-SDA

+12V power

Purple

6.

Thermistor B

Thermistor B

Thermistor B

+24V power

White

7.

GND

Not used

GND

V3 signal

Black

8.

+ 5V

Not used

+ 5V

GND for signal

Green

9.

Cathode electrode

Cathode electrode

Cathode electrode

-5V power

Blue

10.

Not used

Not used

Not used

Temperature

Brown

Measurement alarm relays

The three output relays are located on the measurement board.
They can be individually configured to Normally Open (NO) or to Normally Closed (NC) by
physically moving the jumper on each relay. The illustration is for the EC measurement
board (the positions are different for the TC measurement board):

• Upper relay is set to NC
• Middle relay is set to NO
• Lower relay is shown with no jumper

Note: For all measurement boards, J4 is relay 1, J5 is relay 2 and J6 is relay 3

Sensor installation

EC sensors

For EC sensor installation, servicing, and maintenance ensure you follow the instructions in the
Sensor Installation and Maintenance manual that was supplied with the instrument.

TC sensors

For TC sensor installation, servicing and maintenance ensure you follow the instructions in the TC
Sensor Installation and Maintenance
manual that was supplied with the instrument. Pay particular
attention to the installation and connection of the purge gas supply.

N O T I C E

Do not place the TC sensor into a liquid sample until a constant supply of dry purge gas has been connected, as
liquid could condense inside the measuring chamber and cause damage to the thermal conductor chip.

To ensure the continuation of purge gas while the sensor is in contact with the sample, it is highly
recommended to use a backup purge gas cylinder with an automatic changeover valve that activates
when the first cylinder is empty.
The use of an ORBISPHERE Model 29089 gas regulator (or similar) is also recommended to deliver
a constant, pressure regulated supply of dry purge gas to the sensor, filtered to 40 μm.
In addition, and to prevent any damage to the sensor electronics, the use of a purge safety backup
unit (ORBISPHERE Model 32605) is highly recommended to ensure the supply of purge gas remains
uninterrupted to the sensor in the event of a mains power outage.
The above ORBISPHERE accessories are explained in more detail in the TC Sensor Installation
and Maintenance
manual.

User interface

Instrument controls

The instrument front panel provides:

12 English