Numeric view configuration

1. Select Configure from the View menu followed by Conf. numeric view to customize the display:

Option

Description

Display temperature

Selct Channel temperature to display the sample temperature.

Display mini graph

Check the box to display the graph.

Display time base

Check the box to display the time base.

Upper bound

Adjust graph upper limit.

Lower bound

Adjust graph lower limit.

Time base

Adjust graph time span.

Grid button

Set up the graph to display the x or y axes, grid, or alarm thresholds.

Auto scale update button Automatically set the graph upper and lower bounds to best fit the actual values

displayed.

Clean button

Clear the graph being displayed and restart.

Statistic view

This feature offers statistical data to match with Total Quality Management tools to better analyze
how a process behaves. The statistics are calculated from the data in the measurement file and
values updated each time a new measurement is added.

Diagnostic view

The diagnostic view contains important information but is only really useful for troubleshooting
purposes.

Measurement menu

Instrument configuration

Continuous mode description

Continuous mode is typically used for process measurement.
Continuous mode cycle

• Every 2 sec. measurements are refreshed on the display
• The relays and the analog outputs are updated
• Measurements are continuously stored in memory (volatile and non volatile memory) according to

individual settings

Continuous mode selection

1. Select Config. instrument from the Measurement menu:

Option

Description

Measurement mode

Measurement mode is locked on Continuous
for on line process.

English 15