5 sensor setup menu, Operation – Hach-Lange NO3D sc Nitrate Sensor User Manual

Page 20

Advertising
background image

20

Operation

4.5 Sensor setup menu

SELECT SENSOR (if there is more than one sensor)

CALIBRATE

CAL.CONFIG.

Note: If once a calibration method is chosen the entries will be displayed in the first
submenu of the calibrate menu.

Select SENSOR CODE, MATX1, MATX1 CL-, MATX2, MATX2 CL-, VALUE CORR,
PREVIOUS CAL or FACTORY CAL

-or-

CAL.CONFIG.>SENSOR CODE
DATE

Displays the date of sensor cartridge start up

SENSOR CODE

Display and entry of the sensor code

-or-
CAL.CONFIG.>MATX1

1 point matrix correction (see

section 4.6.3.1 on page 24

)

DATE

Displays the date of the current correction

CONC MEAS 1

Saving the currently measured measurement

SET NO3–N CONC

Entry and display of the NO

3

–N reference value (lab value)

-or-
CAL.CONFIG.>MATX1 CL-

1 point matrix correction with chloride (see

section 4.6.3.2 on page 24

)

DATE

Displays the date of the current correction

CONC MEAS 1

Saving the currently measured measurement

SET CL- CONC

Entry and display of the Cl

-

reference value (lab value)

SET NO3–N CONC

Entry and display of the NO

3

–N reference value (lab value)

-or-
CAL.CONFIG.>MATX2

2 point matrix correction (see

section 4.6.3.3 on page 25

)

CONC MEAS 1

Saving the currently measured measurement

DATE

Displays the date of the current correction

SET NO3–N CONC

Entry and display of the NO

3

–N reference value (lab value)

CONC MEAS 2

Saving the second measured measurement

DATE

Displays the date of the current correction

SET NO3–N CONC

Entry and display of the NO

3

–N reference value (lab value)

-or-

CAL.CONFIG.>MATX2 CL-

2 point matrix correction with chloride (see

section 4.6.3.4 on page 26

)

Note: Carry out the matrix correction only as a standard addition in the sample matrix.

CONC MEAS 1

Saving the currently measured measurement

DATE

Displays the date of the current correction

SET CL- CONC

Entry and display of the Cl

-

reference value (lab value)

SET NO3–N CONC

Entry and display of the NO

3

–N reference value (lab value)

CONC MEAS 2

Saving the second measured measurement

DATE

Displays the date of the current correction

SET CL- CONC

Entry and display of the Cl

-

reference value (lab value)

SET NO3–N CONC

Entry and display of the NO

3

–N reference value (lab value)

Advertising