Troubleshooting

6.3.2 Troubleshooting during calibration

Symptom

Possible cause

Corrective actions

SENSOR CODE

Sensor code entered incorrectly

Using the certificate, check whether the
sensor code was entered correctly. If not
having the certificate, carry out "Factory
Calibration" (Factory cal).

NITRATE

OFFSET

Error in the last nitrate calibration, sensor
cartridge too old, contaminated, faulty

Repeat the calibration.
Use the previous calibration.
Clean or replace the sensor cartridge.

SLOPE

Error in the last nitrate calibration, sensor
cartridge too old, contaminated, faulty

Repeat the calibration.
Use the previous calibration.
Clean or replace the sensor cartridge.

CHLORIDE

OFFSET

Error in the last chloride calibration, sensor
cartridge too old, contaminated, faulty

Repeat the calibration.
Use the previous calibration.
Clean or replace the sensor cartridge.

SLOPE

Error in the last chloride calibration, sensor
cartridge too old, contaminated, faulty

Repeat the calibration.
Use the previous calibration.
Clean or replace the sensor cartridge.