5 sensor setup menu (continued), Operation – Hach-Lange NO3D sc Nitrate Sensor User Manual

Page 21

Advertising
background image

21

Operation

CALIBRATE(continued)

CAL.CONFIG.>
VALUE CORR

Value correction (see

section 4.6.3.5 on page 27

)

Once value correction is complete, the correction data are displayed in the form of
the MATX2

1st NO3–N NO3DSC

Enter the NO

3

–N concentration (NO3D sc) of

the 1st point

1st CL- NO3DSC

Enter the Cl

-

concentration (NO3D sc) of the

1st point

1st NO3–N LAB

Enter the NO

3

–N concentration (lab) of the 1st

point

2nd NO3–N NO3DSC

Enter the NO

3

–N concentration (NO3D sc) of

the second point

2nd CL- NO3DSC

Enter the Cl

-

concentration (NO3D sc) of the

second point

2nd NO3–N LAB

Enter the NO

3

–N concentration (lab) of the

second point

-or-
CAL.CONFIG.>
PREVIOUS CAL

Selection of one of the last four matrix and value corrections performed

-or-
CAL.CONFIG.>
FACTORY CAL

Selection of the preset sensor code

DATE

Date of last calibration

SENSOR CODE

Enter the sensor code

CONFIGURE

EDIT NAME

Entry or editing of the name. Up to 10 alphanumeric characters.

MEAS UNITS

Selection of mg/L or ppm as measurement unit

SET PARAMETER

Selection of NO

3

–N or NO

3

TEMP UNITS

Selection of °C or °F as temperature unit.

TEMP CORR

Entry of the temperature correction (–1.5 to +1.5 °C or –2.7 to +2.7 °F)

RESPONSE TIME

Entry of the response time (30 sec to 300 sec)

LOG SETUP

Selects the interval for the data logging
(deactivate, 1, 2, 5, 10, 15, 30 min, 1, 2, 6 hr).

CL- COMPENSATE.

Selection of the chloride compensation: On, off or fixed value

SET DEFAULTS

Resets the configuration on the factory setting.

4.5 Sensor setup menu (continued)