5 value correction, Operation – Hach-Lange NO3D sc Nitrate Sensor User Manual

Page 27

Advertising
background image

27

Operation

4.6.3.5 Value correction

Value correction offers the option of subsequently correcting a
matrix at two different concentrations.

Take several samples on various days with different concentrations
and perform an analysis of the samples in the laboratory.

Note: The concentrations should be in a concentration range of at least
half a decade.

1. Note the two displayed nitrate and chloride values, which are

measured with the sensor at the time of the sampling.

2. Then note the measured nitrate laboratory reference value.

All 3 values are one correction point.

3. From the values recorded, select the two correction points with

nitrate concentrations as far apart as possible.

4. Select

CAL CONFIG.> CALIBRATE>VALUE CORR

from the menu

and press

ENTER

to confirm.

Enter the 3 noted values:

5. 1. NO3–N NO3D sc: Enter the displayed NO

3

–N value for the

1st correction point. Press

ENTER

to confirm.

6. 1. CL- NO3D sc: Enter the displayed Cl

value for the 1st

correction point. Press

ENTER

to confirm.

7. 1. NO3–N lab: Enter the measured NO

3

–N laboratory

reference value for the 1st correction point. Press

ENTER

to

confirm.

8. 2. NO3–N NO3D sc: Enter the displayed NO

3

–N value for the

2nd correction point. Press

ENTER

to confirm.

9. 2. CL- NO3D sc: Enter the displayed Cl

-

value for the 2nd

correction point. Press

ENTER

to confirm.

10. 2. NO3–N lab: Enter the measured NO

3

–N laboratory

reference value for the 2nd correction point. Press

ENTER

to

confirm.

Value correction is activated and displayed as

MATX2

.

Conc2

Conc1 10

Ч

2

-------------------------------

=

Advertising