4 matx2 cl- correction, Tx2 cl, Operation – Hach-Lange NO3D sc Nitrate Sensor User Manual

Page 26

Advertising
background image

26

Operation

After determining the laboratory reference value, proceed as
follows:

7. Select

SENSOR SETUP>NO3D SC>CALIBRATE>

SET NO3–N CONC

.

8. Enter the laboratory value (reference value) for NO

3

–N and

press

ENTER

to confirm.

9. Select

MEAS CONC 2

and repeat the sequence from Point 4 to

8 for the second value after a concentration change of at least
half a decade.

10. By confirming the second laboratory value, the Matrix2

correction is activated.

4.6.3.4 MATX2 CL- correction

Select

MATX2 CL-

to calibrate the nitrate and chloride sensor

simultaneously at two different measuring points.

1. Select

SENSOR SETUP>NO3D SC>CALIBRATE> CAL CONFIG

.

2. Select

MATX2 CL-

in the selection window and press

ENTER

.

3. Select

CONC MEAS 1

and wait until the NO

3

–N and Cl

-

values

are stable (drift should be <0.03 mg/L).

4. Press

ENTER

to confirm. The value for nitrate and chloride are

saved.

5. Immediately after measuring, take a water sample for

laboratory analysis from as close to the sensor as possible.

6. Analyze the sample immediately after taking the sample as its

nitrate content can change quickly.

After determining the laboratory reference value, proceed as
follows:

1. Select

SENSOR SETUP>NO3D SC>CALIBRATE>SET NO3-N

CONC

.

2. Enter the laboratory value (reference value) for NO

3

-N and Cl

-

press

ENTER

to confirm.

3. Select

MEAS CONC 2

and repeat the sequence from

MEAS

CONC 1

for the second value after a concentration change of at

least half a decade.

4. By confirming the second laboratory value, the Matrix2

correction is activated.

CALIBRATE

CAL. CONFIG.
CONC MEAS 1

DATE

SET CL- CONC

SET NO3–N CONC

CONC MEAS 2

DATE

SET CL- CONC

SET NO3–N CONC