2 menu overview, Menu overview, Operation – Hach-Lange SIGMATAX2 User Manual

Page 30

Advertising
background image

30

Operation

5.2

Menu level 1

Menu level 2

Menu level 3

Description

+SETTINGS

for parameterizing the
instrument

Language

Option to select the language for the menus

HEATER

Self-adjusting pipe trace heater —
turn on if there is a risk of freezing!

Interval

Time interval (6–60 minutes) between
2 samples.
An interval > 12 minutes is only necessary for
longer analysis periods

Probe type

Length of connection tube (10–30 m
(393.7–1181.1 in)

Contrast

Reading angle (0–30) of the display;
temperature dependent

press. indic.

Pressure display on the screen; can be
switched off

Version

Instrument type and program version

+OUTPUT TEST

for testing

connected

data cables

error indication

Switches on the potential-free fault indication
contact for 2 seconds
The fault indication contact is only held in
the previous position for as long as the
user activates it for testing purposes.

fiber optic 1

Blanks fiber optic cable 1 for 1 second

fiber optic 2

Blanks fiber optic cable 2 for 1 second

OF output 2
+SERVICEMENU

for maintenance tasks
and functional tests

+INSPECTION:
SIGMATAX switches to
service mode

clean.samp. tube

Cleans the sample tube (assisted by the
program)

cleaning probe

Cleans the sampling probe (assisted by the
program)

blowing probe

Automatically cleans the sampling probe with
air

blowing valve

Automatically cleans the drain valve with air

+FUNCTIONSTEST

SIGMATAX switches

to service

mode

compressor

Activates compressor for 30 seconds

probe valve

Automatically open and close the

probe valve

ultrasonic

Takes a new sample and activates

ultrasound for 30 seconds

drain valve

Automatically opens and closes the drain
valve

+ERROR STATUS

filling error

Counter for filling errors that have occurred

delivery error

Counter for delivery errors that have occurred

error reset

Reset the counters to "0"

Menu overview