Section 1 specifications – Hach-Lange SIGMATAX2 User Manual

Page 5

Advertising
background image

5

Section 1

Specifications

From software version 1.6. Subject to change.

Measurement

Principle of operation

Samples delivered under pressure, homogenized using ultrasound

Interval

12 to 20 minutes

Sample quantity

For up to 2 PHOSPHAX sigma or TOCTAX process photometers

Tube length

10 m, 20 m, 30 m (393.7 in, 787.4 in, 1181.1 in)

Suction height

Maximum 7 m (275.6 in) (with tube of length 30 m (1181.1 in): maximum 6 m 236.2 in)

Environmental conditions

Ambient temperature

Control unit:

+5 °C to +40 °C (+41 °F to +104 °F)

Connection tube:

–20 °C to +40 °C (–4 °F to +104 °F)

(Without additional insulation; in the event of low external temperatures, correct operation can
only be guaranteed with additional insulation. More information available on request)

Medium temperature

+5 °C to + 30 °C (+41 °F to +86 °F) (probe)

Equipment properties

Protection type

in accordance with IP 54 (Control unit)

Outputs

Fault indication contact, potential-free, 24 V, maximum 1 A

Service interface: RS 232

Power connector

230 V AC / 50 Hz

Power consumption

Device with system components : 600 VA
Recommended fuse: 10 A, time-delay
Line

Heater off

Heater on (–20 °C (–4 °F))
Continuous
operation

At startup

10 m (393.7 in)

150 VA

250 VA

1300 VA

20 m (787.4 in)

150 VA

350 VA

2300 VA

30 m (1181.1 in) 150 VA

470 VA

3000 VA

Inspection interval:

3 months

Maintenance requirement:

Approximately 0.5 h/week, typically

Dimensions:

Open loop control unit: W × H × D 366 Ч 560 Ч 212 mm (14.4 Ч 22 Ч 8.3 in
Probe: Ш Ч L 133 × 404 mm (5.2 × 15.9 in

Earth (approximate)

Control unit:

12 kg

Sampling probe 10 m (393.7 in) 7.5 kg
Sampling probe 20 m (787.4 in) 15 kg
Sampling probe 30 m
(1181.1 in)

22 kg

Mounting poles:

10 kg