Section 6 maintenance – Hach-Lange SIGMATAX2 User Manual

Page 33

Advertising
background image

33

Section 6

Maintenance

Maintenance schedule

DOC273.72.04008.MRZ03

Type of
device:

◦ 10 m

◦ 20 m

◦ 30 m

Instrument number:

Commissioning on:

by:

Annually:
Seal membrane on relief valve

Every 1.5 years:
Replace compressor

Every 3 months:
Clean supply lines and
sampling probe (interior)

Test pressure

Every 6 months:
Replace valve piston with seal
kit

Heat if there is a risk of
freezing!

Weekly:
Clean sample storage vessel

Visually inspect sample quality

Every 1–4 weeks:
Check/clean sampling probe

Clean sample tube
(semi-automatic)

The manufacturer recommends that an inspection contract be concluded. This contract
extends the warranty period to 5 years and makes sure that all inspection and
maintenance tasks are carried out by qualified experts.

Maintenance tasks for users are limited to regular visual inspections and, if necessary,
cleaning procedures.

Advertising