Troubleshooting

As soon as a fault leads to an interruption in operation, SIGMATAX 2 immediately
switches to service mode. Corresponding error messages are shown both on the
SIGMATAX 2 display and on the displays of the connected process photometers.

Service mode means that:

The instrument is no longer in operational mode

The sampling probe and sample storage vessel are empty

The system is depressurized

The sampling probe is closed

The fault indication contact is held in the last valid position

The instrument will not return automatically to operational mode

Error message

Possible cause

Measures

delivery error

Sample line clogged

Select [clean.samp. tube]and carry out a full clean; if
necessary, blow air through the sample line using the
compressor. Inspect the upper level indication sensor.
Check the compressor in the [+FUNCTIONTEST]
menu

filling error

Probe valve clogged

Remove probe, extract blockages and inspect the
probe valve and drain valve.

Sensor error

Internal electrical cause

Call customer service.