Section 7 troubleshooting – Hach-Lange SIGMATAX2 User Manual

Page 37

Advertising
background image

37

Section 7

Troubleshooting

As soon as a fault leads to an interruption in operation, SIGMATAX 2 immediately
switches to service mode. Corresponding error messages are shown both on the
SIGMATAX 2 display and on the displays of the connected process photometers.

Service mode means that:

•

The instrument is no longer in operational mode

•

The sampling probe and sample storage vessel are empty

•

The system is depressurized

•

The sampling probe is closed

•

The fault indication contact is held in the last valid position

•

The instrument will not return automatically to operational mode

Error message

Possible cause

Measures

delivery error

Sample line clogged

Select [clean.samp. tube]and carry out a full clean; if
necessary, blow air through the sample line using the
compressor. Inspect the upper level indication sensor.
Check the compressor in the [+FUNCTIONTEST]
menu

filling error

Probe valve clogged

Remove probe, extract blockages and inspect the
probe valve and drain valve.

Sensor error

Internal electrical cause

Call customer service.