5 access service mode, Operation 5.3.5 access service mode – Hach-Lange SIGMATAX2 User Manual

Page 32

Advertising
background image

32

Operation

5.3.5

Access service mode

1. Press one of the function keys, F1 to F4, for 3 seconds.

2. Press F4 twice to scroll to the [+SERVICEMENU] menu.

3. Press F2 to open this menu.

4. When the [+INSPECTION] or [+FUNCTION TEST] menus are selected, the system

automatically switches to service mode.

Service mode means that:

•

The instrument is no longer in operational mode

•

The sampling probe and sample storage vessel are empty

•

The system is depressurized

•

The sampling probe is closed

•

The fault indication contact remains in the last valid position

•

The instrument will not switch back automatically to operational mode