32

Operation

5.3.5

Access service mode

1. Press one of the function keys, F1 to F4, for 3 seconds.

2. Press F4 twice to scroll to the [+SERVICEMENU] menu.

3. Press F2 to open this menu.

4. When the [+INSPECTION] or [+FUNCTION TEST] menus are selected, the system

automatically switches to service mode.

Service mode means that:

The instrument is no longer in operational mode

The sampling probe and sample storage vessel are empty

The system is depressurized

The sampling probe is closed

The fault indication contact remains in the last valid position

The instrument will not switch back automatically to operational mode